The High Level Pressure Calibration Workstations feature accurate dynamic calibration of pressure sensors over a high pressure range (200 to 15 000 psi) using a stable tourmaline reference sensor. The system is available as Option 9155D-913 for use with a Precision Sensor Calibration Workstation Model 9155D. It is also available as a standalone workstation Model 9913C. A pressure impulse is simultaneously applied to both the reference and sensor under test (SUT) to determine the sensitivity of a pressure sensor at a particular amplitude. By varying the amplitude level, the linearity of the SUT can also be determined. Using a high quality data acquisition system, the system automatically measures, stores and reports calibration results on an ISO 17025 compliant calibration certificate.
Download Datasheet | Download Specification Sheet
9155D-913 Option | |
---|---|
Amplitude Range - psi (MPa) | 200 to 15 000 (1.4 to 103.4) |
Typical Rise Time - mµsec | 3 |
Measurement Technique | Dynamic impulse |
Measurements Supported | Sensitivity, amplitude linearity |
Reference Sensor Type | Tourmaline charge piezoelectric |
Actuator Kit Model | K9913C |
The High Level Pressure Calibration Workstations offers pressure calibration capabilities for existing 9155D Precision Sensor Calibration Workstations. The Option 9155D-913 includes the 9913 Hydraulic Impulse Calibrator, sensor, and software verification to the 9155 base system. The PC system controller and data acquisition hardware are core to the base 9155 workstation, while the -913 control software runs integrated with the 9155 user interface and database. Model K9913C01 upgrades a PCB Model 913B02 to a turnkey K9913C.